| ≥1 piece |
Leak Tester MFY-05 is designed for leak test for food, drugs, cosmetics, industrial products such as flexible packaging, bottles and other packaging materials. MFY-05 adopt negative pressure test mode, automatic chronograph automatic replenishment vacuum packaging, is high-precision automatic leak testing machine. Sealing process and sealing performance can be effectively tested and evaluated.
Test principle
Leak tester using the negative pressure method principle, vacuum the vacuum chamber to make the immersed water sample have internal and external pressure difference. Observe gas escape situation to determine the sealing performance of the sample. Or vacuum the vacuum chamber to make the immersed water sample have internal and external pressure difference. Observation sample expansion and shape of the samples after the release of vacuum recovery and liquid infiltration condition to determine the sealing performance of the sample.
Application
|
Product |
Name |
Application |
|
|
Bags |
Leak Tester is designed for leak test for food, drugs, cosmetics, industrial products such as flexible packaging, bottles and other packaging materials |
|
|
Medical plastic bottle |
Test food, drugs, cosmetics, industrial products such as flexible packaging, bottles and other packaging materials |
|
|
Sterile bottles |
Leak Tester is designed for leak test such as flexible packaging, bottles and other packaging materials |
|
|
Blister package |
It is designed for leak test for food, drugs, cosmetics, industrial products |
?Technical feature
★Leak tester is controlled by micro-computer, digital display vacuum degree
★The instrument utilizes the digital display design of vacuum degree and vacuum retention time to ensure the accuracy of test data
★Top quality parts and components are used to ensure reliable overall product performance
★Specimens could be tested in the preset vacuum condition by automatic compensation of constant pressure
★Automatic end test
★Automatic back flushing
Main Specification
|
Technical parameter |
|
|
Vacuum Degree |
0-90kPa |
|
Vacuum Accuracy |
1FS |
|
Gas Supply Pressure |
0.5MPa~0.7MPa |
|
Vacuum Retension Time |
0~9999.99Hours |
|
Vacuum Room Size |
Φ270 mm x 210 mm (H) (Optional) |
|
Machine size |
430mmX330mmX170mm |
|
Weight |
9Kg |
|
Environment requirements |
|
|
Air Pressure |
0.5-0.9MPa |
|
Environment temperature |
15℃-50℃ |
|
Relative humidity |
No more than 80%,no dewdrop |
|
Power supply |
220V,50Hz |
Standards
ASTM D3078,GB/T 15171,GB/T 27728
Configuration
Standard configurations: Mainframe, Vacuum Chamber
Note: Customers will need to prepare gas supply by himself